1. Introduction
Product Name: colloidal gold test paper reader, colloidal gold immunochromatography reader, lateral flow immunoassay analyzer, gold standard detector, immunoassay instrument, gold standard reader, etc.
Related Products: supporting reagents, usually XX test reagents/test papers (colloidal gold method)
Range of Application: This product is suitable for qualitative/semi-quantitative/quantitative interpretation of sample results by measuring the reflectance of the strips in the reaction zone of the colloidal gold reagent card in a medical laboratory. Other similar products can also be used for rapid immunoassays using fluorescent labels or other labeling methods.
Working Principle: The colloidal gold reader is an instrument for interpreting the results of the colloidal gold reagent card test. The reagent card to be detected is placed in the instrument, and the reflectivity characteristic of the detection reagent card is converted into a photoelectric signal by the sensor, and the photoelectric signal is converted into a corresponding concentration value by the calibration curve information, and the object to be tested is analyzed. Colloidal gold analyzers can be divided into three types according to different light sensors: CCD (Charge Coupled Device), CMOS (Complementary Metal Oxide Semiconductor), and photodiode.
Typical Structure: Generally composed by the host (including photoelectric detection module, mechanical scanning control module, control board module, liquid crystal display module, housing, etc.), information acquisition module (such as: 2D barcode scanner, IC chip reader), power supply, random software, etc.
Product Type:
1) According to the size of the device can be divided into desktop devices, portable (handheld) devices
2) According to the type of power supply, it is divided into internal power supply and network power supply equipment.
3) According to the number of simultaneous tests, it can be divided into multi-channel devices and single-channel devices.
4) According to the test performance can be divided into qualitative, semi-quantitative or quantitative equipment
Desktop multi-channel Desktop single-channel Portable
Product Features:
1) The sample processing method is simple, the reagent usage method is simple, and the instrument operation process is simple.
2) The operating environment is low in requirements and fast in response, and POCT (Point-Of-Care Testing) can be realized.
3) Due to the limitations of reagent technology, the quasi-determination and sensitivity of the product are not as high as those of chemiluminescence.
2. Classification
Classification Code: 22-04-04 Class II immunochromatographic analysis instrument
Product Description: It usually consists of photoelectric detection module, mechanical scanning control module, control board module, information acquisition module and so on. The principle is generally that the reflectance signal of the detection reagent card is converted into a photoelectric signal by a sensor, and the photoelectric signal is converted into a corresponding concentration value or threshold value by the calibration information, and the object to be tested is analyzed.
Expected Usage: Used in conjunction with an adaptation reagent for qualitative and/or quantitative analysis of the analyte in a human sample.
3. Registration Unit
In principle, the registration unit of the colloidal gold reader is based on the technical principle, structural composition, performance index and applicable scope.
1) Different signal acquisition principles should be considered into different registration units, such as CCD, CMOS and photodiode;
2) Different types of electric shock protection should be considered for different registration units;
3) Different degrees of automation should be considered into different registration units;
4) Qualitative, semi-quantitative, quantitative detection functions can be considered to be assigned to the same registration unit.
4. Clinical Trials
Animal Test or Small Sample Test: not required
Clinical Trials: not required
In ‘Medical Device Catalog Exempted from Clinical Trials’
Usually consists of photoelectric detection module, mechanical scanning control module, control board module, information acquisition module and so on. The principle is generally that the reflectance signal of the detection reagent card is converted into a photoelectric signal by a sensor, and the photoelectric signal is converted into a corresponding concentration value or threshold value by the calibration information, and the object to be tested is analyzed. Used in conjunction with an adaptation reagent for qualitative and/or quantitative analysis of the analyte in a human sample.
5. Technical Requirements
Table 1 Relevant Product Standards
Standard Number | Standard Name |
GB/T 191-2008 | Packaging storage and transportation icon |
GB 4793.1-2007 | Safety of electrical equipment for measurement, control and laboratory use - Part 1: General requirements |
GB 4793.9-2013 | Safety of electrical equipment for measurement, control and laboratory use - Part 9: Particular requirements for automatic and semi-automatic equipment for laboratory analysis and other purposes |
GB 4943.1-2011 | Information technology equipment - Safety - Part 1: General requirements |
GB/T 14710-2009 | Medical electrical environment requirements and test methods |
GB/T 18268.1-2010 | Electrical equipment for measurement, control and laboratory use - Electromagnetic compatibility - Part 1: General requirements |
GB/T 18268.26-2010 | Electrical equipment for measurement, control and laboratory use - Electromagnetic compatibility - Part 26: Particular requirements In vitro diagnostic (IVD) medical devices |
GB/T 29791.3-2013 | In vitro diagnostic medical devices - Information provided by manufacturers (indicators) - Part 3: Professional in vitro diagnostic equipment |
YY/T 0316-2016 | Medical device risk management for medical devices |
YY/T 0466.1-2016 | Medical devices - Symbols for labeling, marking and providing information on medical devices - Part 1: General requirements |
YY 0648-2008 | Safety of electrical equipment for measurement, control and laboratory use - Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical devices |
YY/T 1582-2018 | Colloidal gold immunochromatography reader |
Guidelines of Colloidal gold immunochromatography reader registration technical review (draft for comments) |
6. Registration Cost and Duration
1) Product Testing
Test items | Testing duration(90 working days total) | Testing fee(USD) |
Safety Performance | 90 | 4,500 |
EMC | 90 | 3,500 |
Notes:
1. The price of EMC is only the price of one operating mode. If there are multiple operating modes, it needs to be re-priced;
2) Technology Approval
60 working days
Another 60 working days if the 1-year supplement is required
3) Administrative Approval
20 working days
If you would like to get the detailed registration proposal, please contact us via md@cirs-group.com.